CAYLAR, instrumentation scientifique
 
 
Fabricant d'électro-aimants,
de teslamètres RMN,
d'alimentations de puissance
 
 
 
Français
English
 

 
CAYLAR, instrumentation scientifique
 
Fabricant d'électro-aimants, de teslamètres RMN, d'alimentations de puissance

 
 
 
 
 
 
 
 
 

High-stability power supply - 200 V / 60 A / 12 kW

 
 

High stability current source

Available in unipolar or bipolar configurations, our power supplies may be considered for replacement or modernization projects involving installations currently equipped with ER 083, ER083, or ER083CS models, particularly for X-band and Q-band EPR/ESR spectroscopy.

The suitability and compatibility of each solution are assessed on a case-by-case basis according to the exact configuration of the existing system. No universal or plug-and-play compatibility is claimed.


Magnetic field regulation can be performed using a Hall-effect probe or NMR measurement with NMR20, NMR25-1, and NMR25-4 teslameters. Depending on the system configuration and operating conditions, these solutions can provide sub-ppm absolute field regulation.

We also offer adaptable or custom-made probe holders for a wide range of electromagnets, including ER 074-type models. Compatibility is individually validated for each installation.


Important notice: Bruker and the ER 083, ER083, and ER083CS product references are mentioned solely to identify third-party equipment that may be considered for a replacement or modernization assessment. Our power supplies are independently designed and manufactured and are not original Bruker products. Our company is not affiliated with, authorized by, endorsed by, or sponsored by Bruker. All third-party trademarks and product names remain the property of their respective owners.

 
 

Technical specifications

  • Configuration: unipolar or bipolar, with four-quadrant operation in the bipolar version

  • Regulation modes: current or magnetic field

  • Maximum output current: 60 A

  • Maximum output voltage: 200 V

  • Maximum output power: 12 kW

  • Available stability classes: 50 ppm, 10 ppm, 5 ppm, or 1 ppm

  • Analog regulation: ultra-low noise and high stability

  • Control: external analog or internal digital

  • Internal digital control: two 20-bit DACs, one operating over the 100% range and the other over the 1% range, providing an effective signed 26-bit resolution

  • Internal control resolution at 60 A: less than 2 µA, in both unipolar and bipolar versions

  • Digital current readback: ultra-low-noise 24-bit ADC

  • Readback resolution: 10 µA

  • Digital current setpoint increment: 1 mA

  • Field regulation: Hall-effect or NMR-based using NMR20, NMR25-1, and NMR25-4 teslameters

  • Digital interfaces: SCPI commands over RS-232 and Ethernet, Modbus protocol available

  • Analog interfaces: ±10 V readback, ±10 V setpoint input, status signals, interlocks, and analog power-enable control

 

Red: Internal reading from CAYLAR porwer supply / Blue: External independent reading / Green: Internal Temperature


 
 
 

MPU 6U

Rack 6U 19" jusq'à 6kW en monopolaire 2 quadrants ou 3kW en bipolaire 4 quadrants.

MPU 6U
 
 
 

MPU 16U

Baie 16U 19" jusqu'à 24kW en monopolaire 2 quadrants ou 12kW en bipolaire 4 quadrants.

MPU 16U
 
 
 

MPU 24U

Baie 24U 19" jusqu'à 48kW en monopolaire 2 quadrants ou 24kW en bipolaire 4 quadrants.

MPU 24U
 
 
 

Technologies

Current sensorSHUNT or 0-FLUX DCCT
Magnetic field sensorHALL or RMN (option)
Technologywith or without any switched mode power supplies
Control loopAnalogic and digital
Speed ramp control (mA/s)
Thermostatically controlled electronics (1ppm Option)
Setpoint resolution (I, U, B)2 x DAC 20-bit - Down to µV
Resolution measurements (I, U, B)True ADC 24-bit

 

Interfaces and control

Drivers
MODBUS
Ethernet
RS232
External setpoint and analog I, U & B copies

 

Stabilityhttps://wumetiers.wysiup.net/v9/engine9/admin/images/popup/valider.png

Class : 50 ppm, 10 ppm, 1 ppm, sub ppm with NMR Field Regulator

Class1 ppm10 ppm50 ppm
Ripple and noise1 mV eff + 10-6 Vn2 mV eff + 10-5 Vn3 mV eff + 10-4 Vn
High frequency noise
Temperature coefficient<1 ppm / °C<2 ppm / °C<5 ppm / °C
Absolut accuracy
<±0.5 ppm
<±5 ppm<±25 ppm
Long term stability (8 h)
<±0.5 ppm
<±5 ppm

<±25 ppm
NMR Regulation (option)±1 mG / ±0.1 µT (no drift)±1 mG / ±0.1 µT (no drift)
 

Monitoring

  • Reading of the internal temperature.
  • Reading of the temperature of the servo electronics.
  • Reading of the water temperature.
  • Reading of the water flow rate.
  • Reading of the analog safety devices.
 

Analog Protections

  • Ramp speed limit.
  • Overheat detection of all heating elements.
  • Over power.
  • Overvoltage.
  • Overcurrent.
  • Analog voltages.
  • Water flow rate.
  • Leakage detection.
  • Condensation detection.
  • Transistor fault.
  • Quench protection for superconductors.
  • 3 external interlock.

 
 

Ask for information

* Mandatory fields.
 
 
 
 
 
 
 
 
 
 

Regulation*